ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL ST; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1888TC/58
Device Problems Kinked (1339); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/29/2017
Event Type  malfunction  
Event Description
It was reported that a patient presented in operating room for a normal pacemaker battery change-out.Upon interrogation, chronically high right ventricular thresholds and sharp angles on the proximal ends of the leads near the generator were noted.Blood was observed inside of the insulation, and a lead repair kit was added to the proximal end of chronic right ventricular lead.Subsequent testing showed stable diagnostics, and no patient symptoms were reported.
 
Manufacturer Narrative
(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL ST
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6720342
MDR Text Key80249038
Report Number2017865-2017-06656
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1888TC/58
Other Device ID Number05414734501705
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/13/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age33 YR
-
-