• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 10CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 10CM; SCS EXTENSION Back to Search Results
Model Number 3341
Device Problem Fluid/Blood Leak (1250)
Patient Problem Device Overstimulation of Tissue (1991)
Event Type  Injury  
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient ((b)(6)) experienced pocket stimulation.Reportedly, fluid intruded into the header of the extension.As a result, surgical intervention was undertaken on (b)(6) 2017 wherein the extension was explanted and replaced which resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 10CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key6721605
MDR Text Key80306505
Report Number1627487-2017-03388
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2017
Device Model Number3341
Device Lot Number4938503
Other Device ID Number05414734402279
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/12/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age43 YR
-
-