Model Number SC-1132 |
Device Problem
Telemetry Discrepancy (1629)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/19/2017 |
Event Type
malfunction
|
Event Description
|
A report was received that the patient was having difficulty linking the remote control to ipg.The patient will undergo an ipg replacement procedure.
|
|
Manufacturer Narrative
|
Sc-1132 (b)(4): the complaint was confirmed.The antenna coil became dislodged from the lcp frame, and one of the coil ends was fractured at the solder site.The two components have been assembled using rtv adhesive, which is applied at the board vendor.Excessive stress and/or fatigue on the wire where it is soldered to the pin are the result of movement between the coil and the frame.
|
|
Event Description
|
A report was received that the patient was having difficulty linking the remote control to ipg.The patient will undergo an ipg replacement procedure.
|
|
Manufacturer Narrative
|
Additional information was received that the patient underwent an ipg replacement procedure.Device malfunction was suspected by the physician.The patient was doing well postoperatively.
|
|
Event Description
|
A report was received that the patient was having difficulty linking the remote control to ipg.The patient will undergo an ipg replacement procedure.
|
|