• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 2088TC/58
Device Problems Break (1069); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/12/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during initial implant, the physician noticed that there was blood underneath the insulation of the right ventricular lead.The lead was replaced.The patient had no adverse effects and was in good condition post procedure.
 
Event Description
New info received stated that there was an insulation anomaly on the right ventricular lead.
 
Manufacturer Narrative
The damage found was sustained during the surgical procedure.The lead was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6726133
MDR Text Key80493271
Report Number2017865-2017-06816
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2019
Device Model Number2088TC/58
Device Lot NumberA000031581
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age84 YR
-
-