Findings: unit was received with normal operating currents.Also passed self-test, rewind test, basic occlusion test, prime test and displacement test.However, unit received stuck in the motor error loop during bolus/basal delivery and motor position encoder error alarm confirmed in the history file.No unexpected battery out limit alarm noted during testing.The motor was tested outside the device and passed.The motor may have had an intermittent failure that was not detected during our testing.Unit had minor scratched lcd window and cracked reservoir tube lip.
|