MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT266
Device Problem Hole In Material (1293)
Patient Problem No Patient Involvement (2645)
Event Date 06/26/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint rt266 infant dual heated evaqua2 breathing circuits were not returned to fisher & paykel healthcare in (b)(4) for investigation.Our analysis is accordingly based on the information and photographs provided by the healthcare facility, and the results of previous investigations on similar complaints.Results: the photographs provided by the healthcare facility showed that there were multiple holes in the inspiratory and expiratory limb of the circuits.Conclusion: we were unable to determine what may have caused the damage found to the complaint circuits.All rt266 infant dual heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.This suggests that the damage occurred to the circuit after it was released for distribution.Our user instructions that accompany the rt266 infant dual heated evaqua2 breathing circuit state the following: - check all connections are tight before use.- perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- set appropriate ventilator alarms.- do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitisers.
 
Event Description
A distributor in india reported to a fisher & paykel healthcare field representative that holes were found in the tubing of rt266 infant dual heated evaqua2 breathing circuits.The units failed the initial leak test before use on a patient.No patient involvement was reported.
 
Manufacturer Narrative
(b)(4).Method: twelve rt266 infant dual heated evaqua2 breathing circuits were returned to fisher & paykel healthcare in (b)(6) for investigation and were visually inspected.Our investigation is also based on similar complaints received from the same distributor.Results: visual inspection revealed multiple holes in the inspiratory and/or expiratory limbs of eight breathing circuits.No fault was found with the remaining four breathing circuits.The packaging of all returned breathing circuits was damaged.Conclusion: the complaint breathing circuits were caused by abrasion, most likely due to transport or storage conditions.Scanning electron microscopy (sem) was performed on similar complaint devices from the same distributor.Sem revealed that the packaging and circuit tubing had been subjected to an external abrasion source, which supports our current investigation findings.All rt266 infant dual heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.In addition, the packaged breathing circuits are visually inspected for damage before being carefully packed into cartons.This suggests the complaint breathing circuits were damaged after they had been released for distribution.Our user instructions that accompany the rt266 infant dual heated evaqua2 breathing circuit state the following: check all connections are tight before use.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.Do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers.
 
Event Description
A distributor in (b)(6) reported to a fisher & paykel healthcare field representative that holes were found in the tubing of rt266 infant dual heated evaqua2 breathing circuits.The units failed the initial leak test before use on a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
sharon kuruvilla
173 technology dr. suite 100
irvine, CA 92618
8007923912
MDR Report Key6742994
MDR Text Key81132715
Report Number9611451-2017-00673
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT266
Device Catalogue NumberRT266
Device Lot Number2100099837
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-