MEDTRONIC NAVIGATION, INC. (LOUISVILLE) SYSTEM 9735542 15W VISUALASE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Catalog Number 9735542
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/28/2017
Event Type  malfunction  
Manufacturer Narrative
Patient identifier not available from the site.Patient age not available from the site.No parts have been received by the manufacturer for evaluation.Part not returned for analysis.
 
Event Description
A medtronic representative reported that, while in a laser induced thermal therapy (litt), the system was unresponsive to clicks on the software through the mouse.It was reported that the issue occurred following the first ablation.The site pulled the fiber back and were unable to click.The mouse would move in the software.A hard restart of the system restored functionality and the site completed the procedure as intended.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no impact on patient outcome.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 9735542 15W VISUALASE
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key6744709
MDR Text Key81114810
Report Number1723170-2017-03074
Device Sequence Number1
Product Code GEX
UDI-Device Identifier00643169792890
UDI-Public00643169792890
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081656
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 07/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735542
Is the Reporter a Health Professional? No
Date Manufacturer Received06/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/25/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Weight54
-
-