• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET Back to Search Results
Model Number N/A
Device Problems Overheating of Device (1437); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/11/2017
Event Type  malfunction  
Manufacturer Narrative
The complaint is being reported by zimmer biomet as (b)(4).The product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during preparation for the surgery, the product's battery pack of this was deformed.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
The complaint is being reported by zimmer biomet as (b)(4).Review of the device history record for 00515047501, lot number z000006379, identified no relevant deviations or anomalies.Product examination found that the battery pack had been damaged from leaking batteries.The leak had caused battery contacts to become corroded.This complaint is confirmed.The root cause of the reported event could not be specifically determined with the provided information.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key6745394
MDR Text Key81132865
Report Number0001526350-2017-00439
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Model NumberN/A
Device Catalogue Number00515047501
Device Lot NumberZ000006379
Other Device ID Number00889024375161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received09/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-