OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR.; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS
|
Back to Search Results |
|
Model Number A22041A |
Device Problems
Break (1069); Detachment Of Device Component (1104); Component Falling (1105)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 07/04/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect medical device has not yet been returned to the manufacturer for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
|
|
Event Description
|
Olympus was informed that during a transurethral resection of the bladder (turb) procedure, the ceramic insulation at the distal end of the resection sheath broke off and fell inside the patient's bladder.However, no fragment remained inside the patient since it was reportedly retrieved and subsequently disposed of in the sharps container.The intended procedure was successfully completed with another similar device and there was no adverse event or patient injury.
|
|
Manufacturer Narrative
|
The suspect medical device was returned to the manufacturer for evaluation/investigation.The evaluation/investigation confirmed that the ceramic insulation at the distal end of the resection sheath is damaged.A large fragment is broken off.The cause of this damage and the breakage of the ceramic insulation is thermal mechanical stress in combination with wear and tear.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection sheath.It is clearly stated as a warning note in the instructions that impact, fall, shock or similar stress can damage the ceramic insulation at the sheath's distal end and that the instrument must not be used if damaged since otherwise this can cause injuries to the patient and/or user.The user apparently did not follow these instructions since the damage and breakage of the ceramic insulation were caused by thermal mechanical stress.Therefore, this event/incident was attributed to use error and the case will be closed from olympus side with no further actions.However, the event/incident will be recorded for trending and surveillance purposes.In addition, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
|
|