• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE HUMERAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE HUMERAL HEAD Back to Search Results
Device Problem Device Slipped (1584)
Patient Problem Fall (1848)
Event Date 07/05/2017
Event Type  Injury  
Manufacturer Narrative
The contribution of the devices to the experience reported could not be determined as the device was not returned for evaluation.Additionally, the device specific information was not provided, precluding a review of the device history record.
 
Event Description
Index surgery: (b)(6) 2009.Revision of left shoulder components due to loose humeral stem, likely due to glenoid wear after fall 8 weeks post-op.This event report was received through clinical data collection activities.
 
Manufacturer Narrative
The contribution of the devices to the experience reported could not be determined as the device was not returned for evaluation.Additionally, the device specific information was not provided, precluding a review of the device history record.Engineering evaluation noted that the revision reported was likely the result of the patient's fall, which led to wear of the glenoid and eventual loosening of the humeral stem.Wear and loosening are addressed in the product labeling.
 
Event Description
Index surgery: (b)(6) 2009.Revision of left shoulder components due to loose humeral stem, likely due to glenoid wear after fall 8 weeks post-op.This event report was received through clinical data collection activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE HUMERAL HEAD
Type of Device
HUMERAL HEAD
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
graham cuthbert
2320 nw 66th court
gainesville, FL 32653
MDR Report Key6746353
MDR Text Key81116927
Report Number1038671-2017-00519
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age78 YR
Patient Weight66
-
-