• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION CNS-6201A; CENTRAL MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION CNS-6201A; CENTRAL MONITOR SYSTEM Back to Search Results
Model Number CNS-6201A
Device Problem Failure to Power Up (1476)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/11/2016
Event Type  malfunction  
Manufacturer Narrative
The customer states that the central nurses station (cns) failed to boot up after a power loss due to an uninterrupted power supply (ups) failure.After the cns came back up the device would not boot due to a hard drive failure.The hard drive was rebuilt and was verified to be operational.Issue caused by ups failure, not the failure of the device.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer states that the central nurses station (cns) failed to boot up after a power loss due to an uninterrupted power supply (ups) failure.After the cns came back up the device would not boot due to a hard drive failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-6201A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key6749411
MDR Text Key81259375
Report Number8030229-2017-00250
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921114131
UDI-Public4931921114131
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/25/2017,07/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-6201A
Device Catalogue NumberPU-621RA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/25/2017
Distributor Facility Aware Date06/30/2017
Device Age76 MO
Event Location Hospital
Date Report to Manufacturer07/25/2017
Date Manufacturer Received07/25/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/18/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-