• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES SELZACH DRILL BIT Ø17 CANN F/PFNA; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES SELZACH DRILL BIT Ø17 CANN F/PFNA; BIT, DRILL Back to Search Results
Catalog Number 309.600
Device Problems Break (1069); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Reportedly there was no patient involvement.Device is an instrument and is not implanted/explanted.(b)(6).Device is not distributed in the united states, but is similar to device marketed in the usa.Manufacturing location: (b)(4).Supplier: (b)(4).Manufacturing date: october 21, 2009.No non-conformance reports were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.A product investigation was completed: a visual inspection under 5x magnification, functional test, drawing review, and dimensional analysis were performed as part of this investigation.A definitive root cause cannot be identified with the limited complaint details.No new, unique or different patient harms were identified as a result of this evaluation.The design is adequate for its intended use when used and maintained as recommended, it did not contribute to the complaint condition.The device was returned with a portion of the distal cutting flute and the proximal coupling had sheared off from its base.The broken fragment was not returned.Relevant dimensions could not be measured due to the nature of the damage however it is unlikely that the design contributed to the complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that parts were detected as damaged during internal inspection of loaner kits.Reportedly there was no patient or procedure involvement.This report is for drill bit which reportedly did not fit with other parts.When the device was received, it was noted that the proximal coupling post and distal tip were broken; the broken portions were not received.This report is for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL BIT Ø17 CANN F/PFNA
Type of Device
BIT, DRILL
Manufacturer (Section D)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ  CH2545
Manufacturer (Section G)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ   CH2545
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6750464
MDR Text Key81434187
Report Number3000270450-2017-10262
Device Sequence Number1
Product Code HTW
UDI-Device Identifier07611819292895
UDI-Public(01)07611819292895(10)425789
Combination Product (y/n)N
Reporter Country CodeRS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 10/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number309.600
Device Lot Number425789
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received07/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-