The devices were returned and evaluations are complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection was performed and noted the ports had a rough/dry texture and a white-colored appearance.Functional test was performed and no leak or other defect were observed.The reported condition was verified but does not affect the functionality of the device.The cause is most likely related to the end user¿s storage environment.As per the evaluation performed by our supplier, exposure of the product in direct light was the assignable cause for discolored and damage component in the investigation results.Should additional relevant information become available, a supplemental report will be submitted.
|