• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - DOMINICAN REPUBLIC INTRAVIA CONTAINER EMPTY; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - DOMINICAN REPUBLIC INTRAVIA CONTAINER EMPTY; CONTAINER, I.V. Back to Search Results
Catalog Number 2J8002
Device Problems Crack (1135); Degraded (1153)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the injection ports on four (4) units of intravia containers showed signs of deterioration and dryness and cracking.The issue was discovered before use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The devices were returned and evaluations are complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection was performed and noted the ports had a rough/dry texture and a white-colored appearance.Functional test was performed and no leak or other defect were observed.The reported condition was verified but does not affect the functionality of the device.The cause is most likely related to the end user¿s storage environment.As per the evaluation performed by our supplier, exposure of the product in direct light was the assignable cause for discolored and damage component in the investigation results.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAVIA CONTAINER EMPTY
Type of Device
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina san cristobal
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina san cristobal
DR  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6751053
MDR Text Key81298515
Report Number1416980-2017-06224
Device Sequence Number1
Product Code KPE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K964853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2J8002
Device Lot NumberDR16A20073
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-