• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER RADICAL-7; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER RADICAL-7; OXIMETER Back to Search Results
Model Number 25054
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/05/2017
Event Type  malfunction  
Manufacturer Narrative
The returned device was evaluated.During testing the device was able to power on using both ac and battery power.The device was able to obtain readings and alarmed audibly and visually under alarm conditions.When alarming the speaker audio was not crisp, and after a few alarms the devices showed the error "audio speaker fault" and stopped functioning.The failure was isolated to the speaker which was replaced for resolution.Following the replacement, the device functioned as designed.
 
Event Description
The customer reported the device will stop reading when silent alarm button is pressed.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADICAL-7
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
Manufacturer (Section G)
INDUSTRIAL VALLERA DE MEXICALI S.A DE C.V.
calzada del oro, no. 2001
parque industrial palaco
mexicali, baja california 21600
MX   21600
Manufacturer Contact
stefan lissmann
52 discovery
irvine, CA 92618-1604
9492977000
MDR Report Key6753138
MDR Text Key81371465
Report Number2031172-2017-00756
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140188
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25054
Device Catalogue Number9500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received07/05/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-