• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-27
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/22/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(6) 2017 that on (b)(6) 2017, upon removal of the sensor pod from the patient body, the sensor wire had detached.The sensor was inserted at the abdomen on the (b)(6) 2017.No additional event or patient information is available.No product or data were provided for evaluation.The reported detached sensor wire could not be confirmed.A root cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
A sensor was returned for evaluation.A visual inspection was performed and found that the wire is missing from the sensor pod and seal carrier.The reported event of a missing or detached wire was confirmed.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6758631
MDR Text Key81593910
Report Number3004753838-2017-59792
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000019
UDI-Public(01)00386270000019(241)9500-27(10)5222686(17)180113
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/13/2018
Device Model Number9500-27
Device Catalogue NumberSTS-GL-011
Device Lot Number5222686
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
Patient Weight67
-
-