• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY CAPTURA SERRATED LARGE FORCEP-NO SPIKE; FCL, FORCEPS, BIOPSY, NON-ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY CAPTURA SERRATED LARGE FORCEP-NO SPIKE; FCL, FORCEPS, BIOPSY, NON-ELECTRIC Back to Search Results
Catalog Number DBF-2.4SL-230-40-S
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/10/2017
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A review of the device history record could not be conducted because the lot number was not provided.Investigation evaluation: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use states the following in regards to product inspection: "beginning at the handle and moving toward the cups, uncoil the forceps making sure not to stretch the cable.Open and close the cups to verify smooth handle operation and appropriate cup action.Become familiar with the amount of handle movement required to operate the cups.If any irregularities are noted, do not use.Note: exercising the handle while the forceps is coiled may result in damage to the performance characteristics of the forceps." prior to distribution, all captura biopsy forceps are subjected to a visual inspection and functional test to ensure proper workability.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic biopsy procedure, the physician used four (4) cook captura serrated large forcep- no spike.The cable broke and the forceps cannot be controlled by the handle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURA SERRATED LARGE FORCEP-NO SPIKE
Type of Device
FCL, FORCEPS, BIOPSY, NON-ELECTRIC
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key6760354
MDR Text Key82569830
Report Number1037905-2017-00520
Device Sequence Number1
Product Code FCL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 08/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberDBF-2.4SL-230-40-S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/10/2017
Event Location Hospital
Date Manufacturer Received07/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-