The customer contacted the siemens customer care center (ccc).Ccc instructed the customer to run quality control (qc) on both instruments, the results were satisfactory.The ccc specialist instructed the customer to run patient' correlation, resulting satisfactory.The customer stated that she had replaced the sample probe cleaner and the system diluent on the day the event occurred.Ccc instructed the customer to run precision with a known sample, validating results accuracy, for both within cups and between cups, this resulted accurate.The customer examined the peristaltic pump tubing and integrated multi-sensor technology (imt) tubing and cleaned the imt probe.The customer ran bleach and hot water through the sample vent ports, primed all imt fluids, aligned imt probe and module, which passed.The customer ran check 1 on imt probe and home all modules and reset the instrument.The customer replaced the v lyte multi-sensor and ran qc on levels 2 and 3, resulting within range.However, level 1 qc generated measurement errors, after which the customer repeated qc from fresh material in cups, and resulted within ranges.A siemens customer service engineer was dispatched at the customer site.After analyzing the instrument, the cse performed a power flush on imt module and aligned it.The cse found that the connection to the ppd pump was leaking, and tightened the fitting.The cse primed the ppd pump.The cse performed a quick check and ran precision, all resulted within specifications.The cse ran qc on lytes, resulting within range.The cse returned to the customer site for a follow up visit.The cse found the metering pump leaking due to improper mounting, and replaced it.The cse proactively replaced all valves on the imt pump module.The cse primed system fluids and ran quick check; all parameters resulted within range and within the acceptable standard deviation limits.The cse ran all levels of qc, resulting within range.The cause of the discordant, falsely elevated na results is unknown.The instrument is performing within specifications.No further evaluation of the device is required.
|
Discordant, falsely elevated sodium (na) results were obtained on two patient samples on a dimension vista 1500 instrument.The discordant results were not reported to the physician(s).The samples were repeated on an alternate dimension vista instrument, resulting lower.The repeated results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated na results.
|