MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000
Device Problems Electrical /Electronic Property Problem (1198); Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 07/07/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative inspected the imaging system on-site and found the fuses on the mobile view station (mvs) had blown.The fuses were replaced and the issue was resolved.The blown fuses have not been received by the manufacturer for evaluation as they were discarded.A full imaging system check-out was completed following part replacement and all tests passed.Full system functionality was confirmed and the system was returned to service.
 
Event Description
A site representative reported that, monitor on the imaging system would not turn on.There was no delay to procedure as there was no patient present when this issue was observed.No other information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
stephanie riley
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7633987612
MDR Report Key6761892
MDR Text Key81682089
Report Number1723170-2017-03146
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 08/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-