A specific root cause was not identified.Based on the data provided, the investigation did not confirm the issue related to the customer's question as to why no instrument alarm occurred at the time of the sample noise alarm for the low prolactin result.Hitachi was provided with a failure report and concluded that the sample noise alarm was issued when the lh result was sampled.The sample noise alarm did not occur when prolactin was sampled, therefore no data alarm was issued for the prolactin results.Potential root causes for the low prolactin result may be related to the sample adhering to the wall of the tube, the 13 mm sample tube was tilted or the rack adapter for the 13 mm tube was not set.The 13 mm tube may have been tilted and the sample that adhered to the wall of the tube was sampled accordingly.The sample/reagent probe may not have been properly adjusted.
|