MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H802228240022
Device Problems Kinked (1339); Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 06/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was returned for analysis.Visual inspection of the device noted the distal tip to be kinked.Also, the solder which joins the spring distal tip with the core wire was missing.Dimensional inspection of the device was noted to be within specification.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
 
Event Description
Same case as mdr id: 2134265-2017-08108.Reportable based on device analysis completed on 10-july-2017.It was reported that the rotawire was kinked.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending artery.A 330cm rotawire¿ and rotablator burr were selected for use.Upon unpacking, the rotawire was noted to be kinked.Another of the same rotawire was used to complete the procedure.However, device analysis revealed a missing solder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE¿ AND WIRECLIP¿ TORQUER
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6769359
MDR Text Key81942534
Report Number2134265-2017-07386
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729195566
UDI-Public(01)08714729195566(17)20180819(10)19617494
Combination Product (y/n)N
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 07/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/19/2018
Device Model NumberH802228240022
Device Catalogue Number22824-002
Device Lot Number19617494
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-