Model Number E601
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/16/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).
 
Event Description
The customer received questionable high elecsys insulin results for one patient sample.The initial result was 46.65 uu/ml and was reported outside the laboratory.The repeat result was 45.22 uu/ml.The sample was then sent to another laboratory and tested by clia method with a result of 4.2 uu/ml.There was no allegation of an adverse event.The reagent lot number was 157805.The expiration date was requested but was not provided.
 
Manufacturer Narrative
A specific root cause could not be determined.A general reagent issue could be excluded based on the provided qc data.The higher elecsys insulin results compared to competitors were possibly due to anti-insulin antibodies in patients if they were treated with bovine, porcine or human insulin.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 6000 E 601 MODULE
Type of Device
IMMUNOCHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key6770090
MDR Text Key81921170
Report Number1823260-2017-01617
Device Sequence Number0
Product Code CFP
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K060373
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2017
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberE601
Device Catalogue NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age32 YR
-
-