Model Number SC-3400-30 |
Device Problems
Bent (1059); Fracture (1260); High impedance (1291); Kinked (1339); Cut In Material (2454)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 07/17/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional suspect medical device component involved in the event: model#: sc-2316-50 serial #: (b)(4) description: infinion 1x16 perc lead kit50cm model#: sc-3400-30 serial #: (b)(4) description:infinion splitter 2x8 kit30 cm model#: sc-1132 serial #: (b)(4) description: precision spectra implantable pulse generator model#: sc-4316 lot #: 16240019 description: next generation anchor kit-sterile the explanted clik anchor was not returned to bsn as it was discarded by the medical facility.It is indicated that the device will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental med watch will be filed.
|
|
Event Description
|
A report was received that the patient was experiencing inadequate stimulation due to high impedances.The physician suspected that the splitters were corrupted.The patient underwent a revision procedure wherein the ipg, leads, and clik anchor were replaced.During the explant, the leads and splitters were cut around the scar tissue.The patient was doing well postoperatively.
|
|
Manufacturer Narrative
|
Sc-2316-50 (b)(4) device evaluation indicated that the complaint has been confirmed.Visual inspection found that lead was bent/kinked/damaged at the clik anchor site, 1 cm from the set screw mark.The fracture location was 19 cm from the tip of distal end.X-ray inspection confirmed 11cables were fractured (electrodes 1, 2, 4, 6-9, 13-16).There are no exposed cables at the clik site fracture.The fractured cables resulted in the reported high impedances.Sc-2316-50 (b)(4) device evaluation indicated that the complaint has been confirmed.Visual inspection found that lead was bent/kinked/damaged at the clik anchor site, 1 cm from the set screw mark.The fracture location was 24 cm from the tip of distal end.X-ray inspection confirmed all cables were fractured.There are no exposed cables at the clik site fracture.The fractured cables resulted in the reported high impedances.Sc-3400-30 (b)(4) device evaluation indicated that the source of the complaint in regard to the high impedances was verified to be with the associated leads.Visual inspection revealed that the lead splitters were cleanly cut at 21 cm from proximal ends.X-ray inspections found no cable breakage.The clean cut damage is a result of a typical explant procedure and it is not considered a failure.Sc-1132 (b)(4) device evaluation indicated that the complaint in regard to the high impedances was verified to be with the associated leads.The ipg passed all the required tests and revealed normal device characteristics.Sc-4316 lot #: 16240019 it is indicated that the device will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental med watch will be filed.
|
|
Event Description
|
A report was received that the patient was experiencing inadequate stimulation due to high impedances.The physician suspected that the splitters were corrupted.The patient underwent a revision procedure wherein the ipg, leads, and clik anchor were replaced.During the explant, the leads and splitters were cut around the scar tissue.The patient was doing well postoperatively.
|
|
Search Alerts/Recalls
|