The results of the investigation are inconclusive since the device was not returned for analysis.Our investigation was limited to the review of the device history record, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with abbott specifications and procedures.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
During a procedure to repair a mild paravalvular leak in a large defect adjacent to the native valve a 16 mm amplatzer muscular vsd occluder was used (off-label use of the device).The muscvsd was positioned in the defect; however, as the device was being manipulated into position, the device prematurely detached from the delivery cable in the left atrium.The muscvsd was snared and removed.The patient will be re-evaluated for percutaneous closure.
|