• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 4415
Device Problems Device Operates Differently Than Expected (2913); Pumping Problem (3016)
Patient Problem No Patient Involvement (2645)
Event Date 07/14/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The customer was performing cleaning when this issue occurred.The field service representative (fsr) confirmed the reported issue.The cpg pump was not working for speeds 2 and 3, but would work on speed 1.The fsr installed new cpg relay board and completed testing.The unit operated to the manufacturer¿s specifications.During laboratory analysis, the product surveillance technician (pst) tested components on the board with a multimeter and found fuses f5 and f6 to be opened.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during the use of the device for a non-clinical activity, the cardioplegia (cpg) pump was not working on the heater cooler.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6774655
MDR Text Key82602945
Report Number1828100-2017-00351
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 08/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received07/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-