The flash mini device used in the procedure was returned to ostial corporation for evaluation.A two (2) page summary of the evaluation has been attached.No issues were found with the performance of the returned device, though the distal portion of the proximal balloon had clearly been deformed by overexpansion during use.Per the results of the evaluation, it was found that the root cause of the incident was grossly incorrect positioning of the device prior to inflation.Per the results of extensive bench testing, it has been determined that two positioning errors are required for the distal portion of the proximal balloon to significantly expand.The distal portion of the proximal balloon must extend well past the distal edge of a stent being treated and the majority of the working length of the proximal balloon needs to be constrained within the target vessel/stent.These positioning errors result in the proximal portion of the balloon being excessively constrained and can force the inflation volume that would typically fill the proximal portion of the balloon into the distal portion.Per the device ifu, the distal portion of the proximal balloon should not extend past the distal edge of the stent and the proximal portion of the balloon should not be inflated inside of the target vessel.In addition to the case details, ostial corporation reviewed the manufacturing documentation associated with the device lot used during the procedure.No issues were noted that would have contributed to the reported incident.The instructions for use for the product were also reviewed and it was verified that vessel dissection is listed as a potential complication associated with the procedure and that the user is clearly instructed on how the device should be positioned during use.
|
The following information regarding the reported event was obtained from the hospital: following treatment of a ramus intermedius artery, the physician attempted to use the flash mini ostial system in the patient's rca.When inflating the proximal balloon, both ends of the balloon were noted as expanding (dog bone), resulting in a vessel dissection.The patient subsequently coded and expired.
|