TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
|
Back to Search Results |
|
Model Number 816571 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/18/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The fsr cleaned the tube clamp and lubricated the internal components resolving the issue.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the tube clamp was sticky and did not fully close consistently.There was no patient involvement.
|
|