Catalog Number 8065990941
Device Problems Bent (1059); Difficult to Insert (1316)
Patient Problem No Code Available (3191)
Event Date 07/17/2017
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A doctor reported he was unable to complete insertion of the presbyopic corneal inlay of the left eye due to tight tissue opening of the cornea.Upon follow up with company clinical application specialist (cas), it was reported the corneal tissue opening was too tight for the inlay to be inserted.The surgeon opted to abort insertion of the inlay, but did perform a photo refractive keratectomy (prk) enhancement for preparation of the inlay insertion at a later date.Cas also indicated the tool used for inlay insertion was bent making it even more difficult to complete the insertion process.
 
Manufacturer Narrative
The device history records (dhr) for the device was reviewed.No abnormalities that could have contributed to this event were found.The associated device was released based on company acceptance criteria.The system history shows that the laser was verified successfully prior to and after the date of treatment.The root cause could not be identified conclusively.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVELIGHT FS200 FEMTOSECOND LASER
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
nadia bailey
am wolfsmantel 5
erlangen 91058
GM   91058
8176152330
MDR Report Key6787857
MDR Text Key82478623
Report Number3003288808-2017-01855
Device Sequence Number1
Product Code GEX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065990941
Other Device ID Number00380659909412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/09/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age53 YR
Patient Weight68
-
-