Catalog Number 8065990941 |
Device Problems
Bent (1059); Difficult to Insert (1316)
|
Patient Problem
No Code Available (3191)
|
Event Date 07/17/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A doctor reported he was unable to complete insertion of the presbyopic corneal inlay of the left eye due to tight tissue opening of the cornea.Upon follow up with company clinical application specialist (cas), it was reported the corneal tissue opening was too tight for the inlay to be inserted.The surgeon opted to abort insertion of the inlay, but did perform a photo refractive keratectomy (prk) enhancement for preparation of the inlay insertion at a later date.Cas also indicated the tool used for inlay insertion was bent making it even more difficult to complete the insertion process.
|
|
Manufacturer Narrative
|
The device history records (dhr) for the device was reviewed.No abnormalities that could have contributed to this event were found.The associated device was released based on company acceptance criteria.The system history shows that the laser was verified successfully prior to and after the date of treatment.The root cause could not be identified conclusively.(b)(4).
|
|