Dexcom was made aware on (b)(4) 2017, that on (b)(6) 2017, there was an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter.The sensor was inserted at the arm on (b)(6) 2017.Reportedly, the patient used more than one bg meter during their sensor session.No additional event or patient information is available.No data was provided for evaluation.The reported inaccuracy could not be confirmed.A root cause could not be determined.Labeling indicates: do not insert the sensor in sites other than the abdomen (belly) or upper buttocks (for ages 12-17 only).Other sites have not been studied and are not approved.Use in other sites might cause sensor glucose readings to be inaccurate and could result in you missing severe hypoglycemia (low blood glucose) or hyperglycemia (high blood glucose) events.Labeling indicates: finger stick bg values will vary from meter to meter, and test to test.It is important that the patient use the same commercially distributed bg meter during their session.
|