DEXCOM, INC. T:SLIM G4 SYSTEM; CONTINUOUS GLUCOSE MONITOR Back to Search Results
Model Number 9500-27
Device Problems Use of Device Problem (1670); Invalid Sensing (2293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/08/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2017, that on (b)(6) 2017, there was an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter.The sensor was inserted at the arm on (b)(6) 2017.Reportedly, the patient used more than one bg meter during their sensor session.No additional event or patient information is available.No data was provided for evaluation.The reported inaccuracy could not be confirmed.A root cause could not be determined.Labeling indicates: do not insert the sensor in sites other than the abdomen (belly) or upper buttocks (for ages 12-17 only).Other sites have not been studied and are not approved.Use in other sites might cause sensor glucose readings to be inaccurate and could result in you missing severe hypoglycemia (low blood glucose) or hyperglycemia (high blood glucose) events.Labeling indicates: finger stick bg values will vary from meter to meter, and test to test.It is important that the patient use the same commercially distributed bg meter during their session.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM G4 SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6788289
MDR Text Key82897394
Report Number3004753838-2017-41465
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00386270000019
UDI-Public(01)00386270000019(241)9500-27(10)NI(17)NI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 07/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9500-27
Device Catalogue NumberSTS-GL-011
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
Patient Weight77
-
-