MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000-G30
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 07/18/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Testing was not performed as the issue was resolved at the time of the event.No parts were replaced.No parts have been received by manufacturer for evaluation.
 
Event Description
A site representative reported that while outside of a procedure, the imaging system remained in initialization.Please wait for 20 minutes.Rebooting the system resolved the issue.There was no patient present at the time of the issue.
 
Manufacturer Narrative
Correction: coding updated to proper value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key6789973
MDR Text Key82575173
Report Number1723170-2017-03304
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000-G30
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/21/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-