MAUDE Adverse Event Report: B.BRAUN SURGICAL SA NOVOSYN VIOLET 2/0 (3) 70CM HR26 (M); SUTURES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN SURGICAL SA NOVOSYN VIOLET 2/0 (3) 70CM HR26 (M); SUTURES Back to Search Results
Model Number C0068042
Device Problem Knotted (1340)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Manufacturing site evaluation: evaluation on-going.
 
Event Description
Country of complaint: (b)(6).It was reported that the knot does not hold when tied.
 
Manufacturer Narrative
Samples received: 3 unopened pouches.Analysis and results: there are no previous complaints of the same batch.We manufactured and distributed in the market (b)(4) units of this batch.There are no units in our stock.We have received three closed samples.Tightness test to the samples received has been performed and the units are tight.We have tested the knot security control of the samples received and the results are into the current range for this thread and size.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfilled b.Braun surgical requirements.Final conclusion: although the results of the samples received fulfil b.Braun surgical specifications, we take note of this incidence in order to assess if new or additional actions are needed.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Anyway, you will receive a credit note for the units sent for analysis.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVOSYN VIOLET 2/0 (3) 70CM HR26 (M)
Type of Device
SUTURES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
Manufacturer (Section G)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP   08191
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
3145515988
MDR Report Key6791669
MDR Text Key82637124
Report Number3003639970-2017-00422
Device Sequence Number1
Product Code GAM
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
K122734
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/15/2020
Device Model NumberC0068042
Device Catalogue NumberC0068042
Device Lot Number115467
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2017
Is the Reporter a Health Professional? No
Distributor Facility Aware Date08/03/2017
Date Manufacturer Received07/28/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-