Samples received: 3 unopened pouches.Analysis and results: there are no previous complaints of the same batch.We manufactured and distributed in the market (b)(4) units of this batch.There are no units in our stock.We have received three closed samples.Tightness test to the samples received has been performed and the units are tight.We have tested the knot security control of the samples received and the results are into the current range for this thread and size.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfilled b.Braun surgical requirements.Final conclusion: although the results of the samples received fulfil b.Braun surgical specifications, we take note of this incidence in order to assess if new or additional actions are needed.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Anyway, you will receive a credit note for the units sent for analysis.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|