The torque wrench handle was returned for evaluation.Visual inspection found signs of superficial wear.Torque testing found the device to be out of specification.Device was disassembled and it was noted that there seemed to be little signs of lubrication.Ratchet showed signs of abrasive wear.The device is approximately 10 years old.It is suspected the device has been used numerous times/torque cycles.A review of the complaint trend analysis was performed and no systemic trend was identified as a result of the analysis.The root cause cannot positively be determined however it is possible wear, and damage to its components over more than 10 years of uses, resulted in the device exerting more torque than intended.As there has been no systemic trends observed, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|