MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SL FOLEY SWIVEL SILICONE TRICOT 25BX; STATLOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SL FOLEY SWIVEL SILICONE TRICOT 25BX; STATLOCK Back to Search Results
Catalog Number FOL0102
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported the statlock fell off of the tricot material.No patient injury was reported.
 
Manufacturer Narrative
Received 1 foley statlock device in original packaging.The reported event was confirmed with an unknown cause.Per the visual inspection, it was observed that the pad was returned with a slight brownish discoloration, folded, glued to itself and the retainer was detached from the tricot base.An opaque fiber was visualized on the tricot base and base presented no visual damages.Furthermore, the retainer exhibits damages on its inner circle circumference (connection with base) possibly related to a pulling force.Per the functional evaluation, the retainer was reconnected to the tricot base and swiveled 360 degrees, a catheter was clamped and the statlock was turned upside down and remained attached.A great pulling force was needed to detach the retainer from its base.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "the statlock® device is a stabilization device for compatible catheters.Contraindications: known tape or adhesive allergies.Warnings and precautions: do not use the statlock® device where loss of adherence could occur, such as with a confused patient, diaphoretic or nonadherent skin, or when the access device is not monitored daily.Observe universal blood and body fluid precautions and infection control procedures, during application and removal of the statlock® device.Minimize catheter manipulation during application and removal of the statlock® device.Daily maintenance: the statlock® device should be assessed daily and changed when clinically indicated, at least every seven days.If pad becomes soiled, wash with soap/water, saline or hydrogen peroxide.Do not use alcohol or prepackaged bathing systems, which could lead to early lifting.If showering/bathing, cover with plastic wrap or waterproof dressing.Conduct skin assessment prior to application and repeat daily per facility protocol.Use clinical judgment on the removal of the statlock® stabilization device if the patient experiences any fluid shifts that may interfere with skin integrity.Application technique: prep; place foley catheter into retainer.Directional arrow should point towards catheter tip, and balloon inflation arm should be next to the clamp hinge.Close lid, being careful to avoid pinching the catheter.Identify securement site by laying the device retainer on the front of the thigh, leaving 1 inch of catheter slack between insertion site and the statlock® device retainer.After placing the statlock® stabilization device off to the side, cleanse and degrease the securement site with alcohol per hospital policy.Let skin dry.Apply skin protectant, in direction of hair growth, to area larger than securement site.Allow to dry completely (10-15 seconds).Using permanent marker, write initials and date of application on the statlock® device anchor pad.Note: always secure catheter into the statlock® device retainer before applying adhesive pad on skin.Place and peel: align the statlock® stabilization device over securement site leaving 1 inch of catheter slack.Make sure leg is fully extended.While holding the retainer to keep the pad in place, peel away paper backing, one side at a time and place tension-free on skin.Removal technique: disengage; open retainer by pressing release button with thumb, then lift to open.Remove foley catheter from the statlock® device.Dissolve: wipe the edge of the pad using at least 5-6 alcohol pads until a corner lifts.Then continue to stroke undersurface of pad with alcohol to dissolve adhesive pad away from skin.Do not pull or force pad to remove." (b)(4).
 
Event Description
It was reported the statlock fell off of the tricot material.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL FOLEY SWIVEL SILICONE TRICOT 25BX
Type of Device
STATLOCK
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
parque industrial aeropuerto
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
parque industrial aeropuerto
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6796221
MDR Text Key82888539
Report Number1018233-2017-04234
Device Sequence Number1
Product Code EYJ
UDI-Device Identifier00801741076114
UDI-Public(01)00801741076114
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Catalogue NumberFOL0102
Device Lot NumberJUBSF883
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received09/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-