Ous mdr - it was reported that an alert was notet by central monitor at the hospital.When a staff arrived at the patient room, flashing led of both sensing and pacing of the reocor was observed.As it was assumed that pacing might have interrupted, the reocor was exchanged immediately.Since the patient had self-pulse (30 bpm) , the event did not cause any adverse events.
|
Upon receipt, the external pacemaker was inspected.No deviation was found related to the clinical observation during analysis.No indication which might have contributed to the clinical observation was revealed.The visual, mechanical and functional analysis revealed no indication of a material or manufacturing problem.
|