• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problems Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/19/2017
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction of the centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review of kit lot f220 was conducted.There were no non-conformances.This lot met all release requirements.A review of kit lot f220 for the reported complaint issue shows no trends.Trends were reviewed for complaint category, centrifuge bowl leak/break.No trends were detected for this complaint category.The assessment is based on information available at the time of the investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.(b)(4).
 
Event Description
The customer called to report a blood leak due to a centrifuge bowl break.A "big boom" was heard during the purging air process of the treatment.The customer indicated that they had correctly installed the kit.No alarms were reported to have occurred during prime, and the customer was not certain if a blood leak alarm had occurred during the treatment.The patient was reported to have been stable and scheduled to receive another ecp treatment in the afternoon.No product will be returned for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS
bedminster NJ
Manufacturer (Section G)
THERAKOS, INC
10 north high street
suite 300
west chester PA 19380
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key6796697
MDR Text Key83143640
Report Number2523595-2017-00153
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date05/01/2017
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXECP
Device Lot NumberF220
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/02/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient Weight63
-
-