ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 03-2722-9
Device Problems Fluid/Blood Leak (1250); Material Separation (1562)
Patient Problem Blood Loss (2597)
Event Date 07/26/2017
Event Type  Injury  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A biomedical technician (biomed) at the user facility reported that a patient experienced blood loss while undergoing hemodialysis (hd) treatment.The following details were provided.The patient has been undergoing hd therapy three (3) times a week for one (1) month.The blood leak occurred where union and reducer pieces of the bloodline are fused together near the blood pump.It is unknown how much blood was loss.The treated was discontinued and the patient was taken to the hospital for evaluation.The bloodline product was not available to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
The reported complaint was not confirmed as the complaint device was not available for manufacturer evaluation.As such, a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.A records review was performed on the reported lot.An investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances identified during the manufacturing process which could be associated with the reported event.In addition, the batch record review confirmed the labeling, material, and process controls were within specification.The lot passed all release criteria.Review of the batch production record (bpr) did not reveal a probable cause for the customer complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6796731
MDR Text Key82768988
Report Number8030665-2017-00581
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100293
UDI-Public00840861100293
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date02/29/2020
Device Catalogue Number03-2722-9
Device Lot Number17BR01182
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received08/21/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-