Model Number 97714
Device Problems Battery Problem (2885); Charging Problem (2892)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 04/05/2016
Event Type  Injury  
Event Description
Information was received from a healthcare professional of a clinical study regarding a patient.Device interrogation and device data on (b)(6) 2016 determined the spinal cord stimulator battery was discharged.The therapy was suspended on (b)(6) 2016.A device diagnosis of the ins unable to recharge was reported.A clinical diagnosis was not applicable.The event was related to the device or therapy, not related to the implant procedure.A recharge session with the manufacturer had not occurred at the time as the patient was considering explant, however, the patient has recently requested a device replacement.Additional information from the healthcare professional of the clinical study reported that the ins was explanted/replaced on (b)(6) 2017.The event resolved without sequelae.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information from the healthcare professional of the clinical study indicated that it was believed that the device was returned, however, there was no documentation of tracking information at the site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6796917
MDR Text Key82775438
Report Number3004209178-2017-17529
Device Sequence Number1
Product Code GZB
UDI-Device Identifier00643169109483
UDI-Public00643169109483
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2014
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/17/2017
Date Device Manufactured08/27/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age75 YR
Patient Weight97
-
-