Catalog Number 10220
Device Problems Misassembled (1398); Device Misassembled During Manufacturing /Shipping (2912)
Patient Problem No Information (3190)
Event Date 07/21/2017
Event Type  malfunction  
Manufacturer Narrative
Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer reported that during a spectra optia exchange procedure, the operator noticed a misassembly of the roller clamp.The roller clamp from the saline return line was placed on there turn line tubing.Patient information and outcome are not available at this time.The spectra optia exchange set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information.
 
Event Description
The customer declined to provide patient information.
 
Manufacturer Narrative
This report is being filed to provide additional information.Investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Root cause: the cause of this defect was related to a misassembly, where the assembler neglected to follow the appropriate manufacturing operating procedure of the disposable set during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Type of Device
SPECTRA OPTIA EXCHANGE SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
steve kern
10810 w. collins ave
lakewood, CO 80215
3032392246
MDR Report Key6800386
MDR Text Key83119891
Report Number1722028-2017-00328
Device Sequence Number1
Product Code LKN
Combination Product (y/n)N
PMA/PMN Number
K153601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 08/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2019
Device Catalogue Number10220
Device Lot Number03A3128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-