Catalog Number BI70002000 |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/21/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
A medtronic representative inspected the imaging system on-site and could not duplicate the issue.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced and no parts have been returned to the manufacturer for evaluation.
|
|
Event Description
|
A site representative reported that, while in during a spinal fusion case, the imaging system was not initializing when attempting to take a scan.There was a reported indication of x-ray tube arch during reboot.The system was able to shut down and complete the scan.There was a reported delay to the procedure of less than 1 hour due to this issue and no impact on patient outcome.
|
|
Manufacturer Narrative
|
Correction: to include legal statement that was not included in previous supplement.
|
|