• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS UNIV-BATTERY CHARGER II; BATTERY, REPLACEMENT, RECHARGEABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS UNIV-BATTERY CHARGER II; BATTERY, REPLACEMENT, RECHARGEABLE Back to Search Results
Catalog Number 05.001.204
Device Problem Overheating of Device (1437)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device seal was damaged, there was a burnt smell coming from the housing and the warranty seal was broken.The device also failed pretest for cutter insertion.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper handling, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that the universal battery charger device had a burnt plastic odor.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIV-BATTERY CHARGER II
Type of Device
BATTERY, REPLACEMENT, RECHARGEABLE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6803831
MDR Text Key83110591
Report Number8030965-2017-14441
Device Sequence Number1
Product Code MOQ
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 08/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.204
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2017
Date Manufacturer Received08/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-