Biocompatibility testing to iso 10993 was successfully completed on skin-contacting surfaces of the lifevest device as well as the blue(tm) defibrillation gel.Device evaluation of electrode belt sn (b)(4) was completed.The reported problem (gel leak) was confirmed.Upon investigation gel was leaking from the rear 1 therapy electrode.During the incoming functional testing, a 1hz simulated normal sinus rhythm signal was applied to the ecg electrodes.The pulse delivery circuitry test verified proper delivery of full energy 150j biphasic pulse.The functional testing confirmed proper ecg acquisition and pulse delivery functionality.The root cause of the gel leak was unable to be positively identified.
|