Model Number ADVIA CENTAUR XP |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/25/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer contacted the siemens customer care center (ccc) after obtaining erratic ee2 results on quality controls (qc) on following daily cleaning procedure (dcp) performed on the instrument.A siemens customer service engineer (cse) was dispatched to the customer's site.After inspecting the instrument, the cse found a leak in the wash manifold valve drawer coming from the fitting tube with wash 1 cleaning solution.The cse cleaned up the liquid and repaired the tube fitting.The cse calibrated ee2 with a new pack of reagent and then ran qc along with the current one on board that generated low relative light unit (rlu), the results were satisfactory except for qc level 1 which obtained low rlu on the new pack.The customer requested the technical applications specialist (tas) to run master curve material (mcms) and run precision to validate these tests.The tas performed accuracy, precision and linearity on the ee2 assay, resulting acceptable.The cause of the discordant, falsely elevated ee2 results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
|
Event Description
|
Discordant, falsely elevated enhanced estradiol (ee2) results were obtained on patient samples on an advia centaur xp instrument.The initial results were reported to the physician(s).The samples were repeated on the same instrument, resulting lower.The samples were then repeated on an alternate advia centaur instrument, and the results matched the original instrument repeat results.The corrected results were reported to the physician(s).There are no reports of patient intervention or adverse health consequence due to the discordant, falsely elevated ee2 results.
|
|
Manufacturer Narrative
|
The initial mdr 2432235-2017-00482 was filed on 21-august-2017.A siemens customer service engineer (cse) had returned at the customer's site on (b)(6) 2017: the cse installed a new separation manifold and all 4 aspirate probe pinch valves.The cse ran quality control (qc), resulting acceptable.Additional information (13-november-2017): a siemens headquarter support center (hsc) specialist evaluated the service information and determined that the wash manifold is essential for the accurate washing of cuvettes and incomplete washing due to a leak, can potentially lead to high discordant results as seen in this event.The cse performed the appropriate corrective action to resolve this issue.The cause of the discordant, falsely elevated ee2 results could not be determined.
|
|
Search Alerts/Recalls
|