Model Number 5826
Device Problem Programming Issue (3014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2017
Event Type  Injury  
Event Description
It was reported that despite multiple attempts the device was not able to be programmed.The device was explanted and replaced.Patient was stable before, during, and after the procedure.
 
Manufacturer Narrative
The reported field event of inability to program the device was not confirmed in the laboratory.The device was tested on the bench and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR XL DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6808661
MDR Text Key83190456
Report Number2017865-2017-08406
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2010
Device Model Number5826
Device Lot Number2703997
Other Device ID Number05414734501620
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age73 YR
-
-