The customer contacted the siemens technical support center (tsc) to report the discordant sodium result.The tsc took system into diagnostics mode and primed integrated multisensor technology (imt) probe, pump, and cleaner.The tsc auto-aligned imt and imt module, which passed without error.The tsc homed modules, exited diagnostics and reset the dimension vista.The tsc had replaced the v-lyte multi-sensor on the system followed by conditioning the sensor and repeated lytes quality control (qc) but it was out of range.The customer performed a dilution check and accepted the new correction factor.Qc was repeated and recovered within the customer's range.The cause of the discordant sodium result is unknown.This instrument is performing according to specifications.No further evaluation of this device is required.
|
A discordant, falsely high sodium result was obtained on one patient sample on a dimension vista 500 instrument.The initial result was reported to the physician(s).The same sample was repeated on an alternate instrument, and recovered lower.The corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely high sodium result.
|