Model Number N/A
Device Problem Sparking (2595)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/05/2016
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Investigation summary: review of the device history record could not be performed as there was no lot number reported for this complaint.Product examination found that the battery pack had ruptured due to a short circuit.However, the returned product included only the battery pack, which had been separated from the hand unit by cutting the wire.Cutting the wire can directly result in the kind of damage both exhibited in the product examination and reported by the customer.The timing of this rupture cannot be determined and this reported event cannot be confirmed.The root cause of the reported event could not be specifically determined with the provided information.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Event Description
It was reported that during surgery a spark came from the battery pack.Subsequently, the nurse opened an alternative one to complete the surgery.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key6813202
MDR Text Key83489724
Report Number0001526350-2017-00566
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 08/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00515047501
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/21/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-