The cycler was returned for evaluation.Visual inspection of the returned cycler exterior showed no sign of physical damage.There are visual indications of dried fluid within the cassette compartment.No burrs or sharps in cassette area that may have punctured a cassette membrane.During the ¿set-up¿ phase of the simulated treatment an "motor pump a¿ alarm occurred three times and the treatment was canceled.Passed mushroom head check.When removing the heater tray a burnt odor was present.An internal inspection of the cycler found c76 (capacitor) on the i/o board was heat damaged.C76 is connect to output pin 3 of ic35 (voltage regulator) which provides +12 volts.A measurement of ic35 output pin 3 showed.09 volts.¿patient sensors 1 and 2¿ requires 12 volts to function.A review of ¿patient sensors 1 and 2¿ in ¿diagnostics mode¿ show sensor 1 displayed reading of (-314/-4.5) and sensor 2 (-308/-4.4).In addition, there was visual evidence of dried fluid under pump ¿a¿ and ¿b¿ mushroom head and within the recess of the bottom cover adjacent to the pump.The cause of the observed dried fluid could not be determined.The reported symptoms were confirmed.An investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the device manufacturing review confirmed the labeling, material, and process controls were within specification.
|