• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problems Crack (1135); No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 08/04/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Initial contact telephone number: (b)(6).Livanova (b)(4) manufactures the s5 roller pump.The event occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative confirmed the reported issue and found that the display illuminates orange without activation.The pump display was replaced.An nvmem clear and functional checks were carried out without further issues.If any additional information pertinent to the reported event is received, it will be submitted in a supplemental report.Device not returned.
 
Event Description
Livanova deutschland received a report that the touch screen of the s5 roller pump was found to have a crack during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada 80004
2812287260
MDR Report Key6825752
MDR Text Key83974442
Report Number9611109-2017-00673
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 08/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/04/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/29/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-