• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM Back to Search Results
Catalog Number 8065990739
Device Problem Burst Container or Vessel (1074)
Patient Problem No Information (3190)
Event Date 08/04/2017
Event Type  malfunction  
Manufacturer Narrative
The device history records (dhr) for the device were reviewed.The associated device was released based on acceptance criteria.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.No udi required due to this device was out of production prior to the september 24, 2014 udi regulation date.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that the nitrogen was spiking up to 600 psi causing the nitrogen line to burst.Additional information requested.
 
Manufacturer Narrative
During a onsite visit the fse (field service engineer) found system had a faulty nitrogen regulator.To fix this issue, the fse replaced nitrogen regulator and repaired the nitrogen line.The root cause could not be identified conclusively.Most possible root cause could be a faulty nitrogen regulator.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
nadia bailey
am wolfsmantel 5
erlangen 91058
GM   91058
8176152330
MDR Report Key6827508
MDR Text Key84085840
Report Number3003288808-2017-01966
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065990739
Other Device ID NumberV4-1.02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/14/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-