• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problems Break (1069); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2017
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction drive tube leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot d361 was conducted.There were no non-conformances.This lot met all release requirements.A review of kit lot d361 for the reported issue shows no trends.Trends were reviewed for complaint category, drive tube leak/break.No trends were detected for this complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.(b)(4).
 
Event Description
Customer called to report a drive tube break during procedure.Customer stated there was about 1250 ml of whole blood processed when a noise was heard and the break had occurred.Customer mentioned that the inside of the centrifuge chamber was covered in blood, however the bowl does not appear to have been shattered.The patient is said to be in stable condition and was not affected by the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS
bedminster NJ
Manufacturer (Section G)
THERAKOS, INC
10 north high street
suite 300
west chester PA 19380
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key6831187
MDR Text Key85613263
Report Number2523595-2017-00162
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)D361(17)171001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date10/01/2017
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXUSA
Device Lot NumberD361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-