DEPUY SYNTHES POWER TOOLS 8 CM SHORT,QD ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT
|
Back to Search Results |
|
Catalog Number QD8-G1 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/18/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Reporter¿s complete address and phone number were not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the bearing on the attachment device was failing.It was not reported if there were any delays in the surgical procedure or whether a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device bearing was failing could not be confirmed.However, it was determined that the device failed the cutter engagement assessment and had foreign debris inside of the bushings which was preventing the cutter from being inserted into the device.The assignable root cause was determined to be due to cleaning sterilization, and /or maintenance procedures not followed per the directions for use.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|