BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) BLAZER PRIME¿ XP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number M004EPTP4500THN40 |
Device Problem
Kinked (1339)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/26/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Device evaluated by manufacturer: the device was returned for analysis.Visual inspection of the device revealed that the shaft is bent at 90 degrees at approximately 1.8cm from the distal tip between ring 2 and ring 3.An end edge of the center support is poking out of the tubing at the bend.Both the ring 2 and ring 3 seals are compromised.Continuity checks revealed no electrical opens or shorts and all electrode, thermistor and sensor resistances measured in specification and were typical.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Tip motion was evaluated against the specified curve template/fixture and both the right and left curve failed to reach the specified template area.X-rays confirmed that the center support is broken and an end is poking up through the catheter shaft tubing.The distal section was dissected and body fluids were noted in the interior of the sheath.The kevlar wrap is slightly displaced; it does not cover the small curves¿ steering wire solder joint on the center support.The steering wires are attached to the center support.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause has been determined to be related to a supplier manufacturing issue.(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on 15-aug-2017.It was reported that a catheter distal end kink occurred.A blazer prime¿ xp was selected for use.During the procedure, a number of ablation burns were already performed when it was noticed under fluoroscopy that the catheter shape appeared to be kinked.Subsequently, the catheter was withdrawn from the patient¿s body and it was confirmed that the distal end of the catheter became kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However device analysis revealed that the end edge of the center support is poking out of the tubing at the bend.
|
|
Search Alerts/Recalls
|
|
|